2:27

Peter Antevy explains when and how to run control tests on glucometers: insert a strip, apply control solution, and check that results fall within acceptable ranges (Control 1: 115-173; Control 2: 208-312). Test when first using a glucometer, opening new strips, seeing error messages, or after dropping the device. Control solution expires 90 days after opening.

Questions this video answers

115–173

208–312

90 days after opening