Problems with a syringe pump II (the sequel)

In this short talk, the speaker examines a number‑entry interface used in a medical syringe pump. The UI features four arrow buttons: two for increasing a value and two for decreasing it. While the small up and down chevrons would be expected to act as exact inverses, the speaker demonstrates that the step sizes are inconsistent—pressing up at a value of 9 increments by 0.1, but pressing down at the same point decrements by 0.3; at higher values (e.g., 9.57) the up button still adds 0.1 while the down button subtracts 0.8. The speaker argues this irregular behavior is likely an unintended design flaw. He suggests that because users cannot predict the effect of each press, the interface forces them to keep their eyes on the screen, and proposes an experiment to measure how much visual attention the UI demands.